Konica Minolta CRT Color Analyzer-CA-100Plus


Specifications

Principal Specifications CA-100Plus

 

CA-100Plus (Measuring Probe)

CA-100Plus (High Luminance Measuring Probe)

Detector

Silicon photocell

Silicon photocell

Measurement diameter

Ø 21 mm

Ø 21 mm

Display range

Luminance

0.01 to 1000 cd/m2

0.01 to 2000cd/m2

Chromaticity

Displayed in 4- or 3-digit values (selectable)

Displayed in 4- or 3-digit values (selectable)

Luminance

Measurement range

0.05 to 1000 cd/m2

0.05 to 2000 cd/m2

Accuracy (for white)*1

±2%, ±1 digit of reading (Temperature: 23°C±2°C, Relative humidity: (40±10)%)

±2%, ±1 digit of reading (Temperature: 23°C±2°C, Relative humidity: (40±10)%)

Repeatability*1

0.05 to 1000 cd/m2
0.2%+1 digit (2σ)

0.05 to 2000 cd/m2
0.2%+1 digit (2σ)

Chromaticity

Measurement range

0.05 to 1000 cd/m2

0.05 to 2000 cd/m2

Accuracy*1

0.05 to 0.19 cd/m2 ±0.006 for white; 0.20 to 0.49 cd/m2 ±0.004 for white; 0.50 to 1000 cd/m2 ±0.003 for white; 40.00 cd/m2 ±0.002 for white, ±0.004 for monochrome*2

0.05 to 0.09 cd/m2 ±0.008 for white; 0.1 to 0.39 cd/m2 ±0.006 for white; 0.40 to 0.99 cd/m2 ±0.004 for white; 1.00 to 2000 cd/m2 ±0.003 for white; 40.00 cd/m2 ±0.002 for white, ±0.004 for monochrome*2

Repeatability*1

0.05 to 0.19 cd/m2 0.006 (2σ); 0.20 to 0.49 cd/m2 0.002 (2σ); 0.50 to 1000 cd/m2 0.001 (2σ)

0.05 to 0.09 cd/m2 0.009 (2σ); 0.1 to 0.39 cd/m2 0.006 (2σ); 0.40 to 0.99 cd/m2 0.002 (2σ); 1.00 to 2000 cd/m2 0.001(2σ)

Measurement speed*3

xyLv

0.05 to 0.99 cd/m2 5 (4.5) meas./sec.*4 ; 1.00 to 1000 cd/m2 20 (17) meas./sec.*4

0.05 to 1.99 cd/m2 5 (4.5) meas./sec.*4 ; 2 to 2000 cd/m2 20 (17) meas./sec.*4

Display

Digital

xyLv, TΔuvLv, RGB, XYZ, u'v'Lv

xyLv, TΔuvLv, RGB, XYZ, u'v'Lv

Analog

Δ x, Δ y, Δ Lv, R/G, B/G, Δ G, Δ R, B/R, G/R

SYNC mode

NTSC, PAL, EXT, UNIV, INT

NTSC, PAL, EXT, UNIV, INT

Object under measurement

Vertical synchronous frequency: 40 to 200Hz

Vertical synchronous frequency: 40 to 200Hz

Memory channels

100

100

Analyzer function

Standard function

Standard function

Interface

USB (Version 1.1) or RS-232C (38,400bps or below)

USB (Version 1.1) or RS-232C (38,400bps or below)

Multi-point measurement

Max. 5 points (Use 4-Probe Expansion Board CA-B04.)

Max. 5 points (Use 4-Probe Expansion Board CA-B04.)

Operating temperature/humidity range

Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation Luminance change: ±2% ±1 digit of reading for white Chromaticity change: ±0.002 for white, 0.006 for monochrome (from reading of Konica Minolta's standard CRT*1, 40.00cd/m2, at 23°C, 40%

Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation Luminance change: ±2% ±1 digit of reading for white Chromaticity change: ±0.002 for white, 0.006 for monochrome (from reading of Konica Minolta's standard CRT*1, 40.00cd/m2, at 23°C, 40%

Storage temperature/humidity range

0 to 28°C: Relative humidity 70% or less with no condensation 28 to 40°C: Relative humidity 40% or less with no condensation

0 to 28°C: Relative humidity 70% or less with no condensation 28 to 40°C: Relative humidity 40% or less with no condensation

Input voltage range

100 to 240V~, 50 to 60Hz, 50VA

100 to 240V~, 50 to 60Hz, 50VA

Dimensions (W × H × D)

Main unit, Probe

340 × 127 × 216 mm, Ø 45 × 142 mm

340 × 127 × 216 mm, Ø 45 × 142 mm

Weight

Main unit, Probe

3.58 kg, 285 g

3.58 kg, 285 g

 


PDF Dosyasını İndir