Konica Minolta Display Color Analyzer-CA-210


Specifications

Principal Specifications CA-210

Receptor

Silicon photocell

Measurement area

Ø 27mm

Acceptance angle

±2.5°; Pointing function: By LED

Measurement distance

30±10 mm

Display range

Luminance

0.01 to 1000 cd/m2

Chromaticity

Displayed in 4- or 3-digit value (selectable)

Luminance

Measurement range

0.10 to 1000 cd/m2

Accuracy (for white) *1

±2%, ±1 digit of reading (Temperature: 23°C ±2°C; Relative humidity:(40±10)%)

Repeatability (2σ) *1

0.10 to 0.99 cd/m2: 0.2%+1digit; 1.0 to 1000 cd/m2: 0.1%+1digit

Chromaticity

Measurement range

0.10 to 1000 cd/m2

Accuracy *1

0.10 to 4.99 cd/m2: ±0.008 for white
5.00 to 39.99 cd/m2: ±0.005 for white
40.00 to 1000 cd/m2: ±0.003 for white
160 cd/m2: ±0.002 for white; ± 0.004 for monochrome *2

Repeatability (2σ) *1

0.10 to 0.19 cd/m2: 0.015
0.20 to 0.49 cd/m2: 0.008
0.50 to 1.99 cd/m2: 0.003
2.00 to 1000 cd/m2: 0.001

Measurement speed (meas./sec.)

xyLv*3

0.10 to 1.99 cd/m2: 5 (4.5)
4 cd/m2 or greater: 20 (17)

Display

Digital

xyLv, XYZ, TΔuvLv, u'v'Lv, RGB analyze

Analog

ΔxΔyΔLv, R/G B/G ΔG, ΔR B/R G/R

LCD

16 characters × 2 lines (with backlight)

SYNC mode

NTSC, PAL, EXT, UNIV, INT

Measurement subject

Vertical synchronous frequency: 40 to 200Hz

Memory channels

100

Analyzer function

Standard function

Interface

USB (Rev. 1.1) or RS-232c (38,400 bps maximum)

Multi-point measurement Max.

5 points (using optional 4-Probe Expansion Board CA-B14)

Software Solutions

SDK software (supplied as standard accessory)

Operating temperature/humidity range

Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation. Luminance change: ±2% ±1 digit of reading for white; Chromaticity change: ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD *1, 160.0 cd/m2 at 23°C 40%RH

Storage temperature/humidity range

0 to 28°C: Relative humidity 70% or less with no condensation; 28 to 40°C: Relative humidity 40% or less with no condensation

Input voltage range

100 to 240V~50/60Hz, 50VA

Dimensions (W × H × D)

Main body: 340 × 127 × 216 mm
Probe: Ø 49 × 204 mm

Weight

Main body: 3.58 kg; Probe: 520 g

 


PDF Dosyasını İndir