Konica Minolta Spectrophotometer Portable-CM-2600d_2500d
Specifications
Principal Specifications CM-2600d
Model
CM-2600d
CM-2500d
Illumination / Viewing system
d:8° (diffuse illumination, 8-degree viewing ), equipped with simultaneous measurement of SCI (di:8° specular component included) / SCE (de:8° specular component excluded) Conforms to CIE No.15,ISO 7724/1,ASTM E1164,DIN 5033 Teil7 and JIS Z8722 Condition C standard.
Sphere Size
Ø 52 mm
Light-receiving element
Silicon photodiode array (dual 40 elements)
Spectral separation device
Diffraction grating
Wavelength range
360 nm to 740 nm
Wavelength pitch
10 nm
Half bandwidth
Approx. 10 nm
Reflectance range
0 to 175%, resolution: 0.01%
Light source
3 pulsed xenon lamps
2 pulsed xenon lamps
Measurement time
Approx. 1.5 seconds (approx. 2 seconds for fluorescent measurement)
Approx. 1.5 seconds
Minim um interval between measurements
3 seconds for SCI/SCE (4 seconds for fluorescent measurement)
3 seconds for SCI/SCE
Battery perfomance
Alkaline manganese:approx. 1000 measurements
Measurement/illumination area
MAV : Ø 8 mm / Ø 11 mm SAV : Ø 3mm / Ø 6 mm (Selectable between MAV and SAV)
MAV : Ø 8 mm / Ø 11 mm
Repeatability
Spectral Reflectance:Standard deviation within 0.1% (360 to 380 nm within 0.2%) Colorimetric Value : Standard deviation within ΔE*ab 0.04 (Measurement conditions:White calibration plate measured 30 times at 10-second intervals after white calibration was performed)
Inter instrument agreement
withinΔE*ab 0.2 (MAV/SCI) Average for 12BCRA Series II color tiles compared to values measured with master body.
UV adjustment
Instantaneous numerical adjustment (no mechanical adjustment required) With UV400nm cut filter
No UV adjustment
Measurement mode
Single measurement/automatic averaging of multiple measurements(auto mode: 3, 5, 8 times/manual mode)
Interface
RS-232C standard
Observer condition
CIE: 2° and 10° standard observer
Illuminant condition
CIE: A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (simultaneous evaluation is possible using two light sources)
Display data
Spectral value/graph, colorimetric value, color difference value/graph, PASS/FAIL result
Color space / colorimetric data
L*a*b*, L*C*h, CMC (1:1), CMC (2:1), CIE94, Hunter Lab, Yxy, Munsell, XYZ, MI, WI (ASTM E313), YI (ASTM E313/ASTM D1925), ISO Brightness (ISO 2470), Density statusA /T, WI/Tint (CIE/Ganz), CIE00
Data memory
1700 sets of data (as SCI/SCE 1 data) * 700 sets of data in the "defined in COND." mode. * Total of the sample data for the COND and TASK modes and color diference target data
Tolerance judgment
Tolerance for color difference (both box and eliptical tolerances can be set)
Power source
4 AA-size battery or AC adapter AC-A305
Dimensions (W × H × D)
69 × 96 × 193 mm
Weight
Approx. 670 g (without batteries)
Operating temperature / humidity range (*1)
5 - 40°C, relative humidity 80% or less (at 35°C) with no condensation
Storage temperature/humidity range
0 - 45°C, relative humidity 80% or (at 35°C) with no condensation
Standard accessories
White calibration plate, Target mask Ø 8 mm, Target mask Ø 3 mm , RS-232C cable, AC adapter,AA-size battery (×4)
Optional Accessories
Hard case, Dust cover set, Dust cover, SpectraMagic(tm)NX (software), Zero calibration box.