Işık Ölçüm Cihazi
Konica Minolta Luminance Meter-LS-100 & LS-110
Specifications
Principal Specifications LS-100/LS-110
Model |
Luminance Meter LS-100 |
Luminance Meter LS-110 |
Type |
SLR spot luminance meter for measuring light-source and surface brightness |
|
Acceptance angle |
1° |
1/3° |
Optical system |
85mm f/2.8 lens; SLR viewing system; flare factor less than 1.5% |
|
Angle of view |
9° |
|
Focusing distance |
1014mm (40 in.) to infinity |
|
Minimum measuring area |
Ø 14.4 mm |
Ø 4.8 mm |
Receptor |
Silicon photocell |
|
Relative Spectral Response* |
Within 8% (f1´) of the CIE spectral luminous efficiency V (λ) |
|
Response time |
FAST: Sampling time: 0.1s, time to display: 0.8 to 1.0s; SLOW: Sampling time: 0.4s, time to display: 1.4 to 1.6s |
|
Luminance units |
cd/m2 or fL (switchable) |
|
Measuring range |
FAST: 0.001 to 299,900 cd/m2 (0.001 to 87,530fL) |
FAST: 0.01 to 999,900 cd/m2 (0.01 to 291,800fL) |
Accuracy |
0.001 to 0.999 cd/m2 (or fL): ±2% ±2 digits of displayed value |
0.01 to 9.99 cd/m2 (or fL): ±2% ±2 digits of displayed value |
(Illuminant A measured at ambient temperature of 20 to 30°C/68 to 86°F) |
||
Repeatability |
0.001 to 0.999 cd/m2 (or fL): ±0.2% ±2 digits of displayed value |
0.01 to 9.99 cd/m2 (or fL): ±0.2% ±2 digits of displayed value |
(Measurement subject: Illuminant A) |
||
Temperature/humidity drift |
Within ±3% ±1 digit (of value displayed at 20°C/68°F) within operatingtemperature /humidity range |
|
Calibration mode |
Minolta standard/user-selected standard (switchable) |
|
Color correction factor |
Set by numerical input; range: 0.001 to 9.999 |
|
Reference luminance |
1; set by measurement or numerical input |
|
Measurement modes |
Luminance; luminance ratio; peak luminance or luminance ratio |
|
Display |
External: 4-digit LCD with additional indications |
|
Viewfinder: 4-digit LCD with LED backlight |
||
Data communication |
RS-232C; baud rate: 4800bps |
|
External control |
Measurement process can be started by external device connected to data output terminal |
|
Power source |
One 9V battery |
|
Power consumption |
While measuring button is pressed and viewfinder display is lit: 16 mA average |
|
Operating temperature/humidity range |
0 to 40°C, relative humidity 85% or less (at 35°C) with no condensation |
|
Storage temperature /humidity range |
-20 to 55°C, relative humidity 85% or less (at 35°C) with no condensation |
|
Dimensions |
79×208×150 mm (3-1/8×8-3/16×5-7/8 in.) |
|
Weight |
850 g (30 oz.) without battery |
|
Standard accessories |
Lens cap; Eyepiece cap; ND eyepiece filter; 9V battery; Case |
Etiketler : Ankyralab KONICA MINOLTA Işık Ölçüm Cihazi Konica Minolta Luminance Meter-LS-100 & LS-110, Laboratory Devices, Laboratory Equipments, Laboratory Tools, Ankyra Laboratory Tools, Ankyralab Laboratory Devices, Ankyralab Laboratory Systems