Specifications
Principal Specifications CS-2000A
Model |
CS-2000A |
||
Measurement angle |
0.1°, 0.2°, 1° selectable |
||
Measurement wavelength range |
380 to 780nm |
||
Wavelenght resolution |
0.9nm/pixel |
||
Display wavelength pitch |
1.0nm |
||
Median wavelength precision |
±0.3nm (at 435.8nm, 546.1nm, and 643.8nm) |
||
Half bandwidth |
5nm or less |
||
Luminance measurment range (for Standard illuminant A) |
0.1° measurement angle: 0.05 to 500,000 cd/m2 0.2° measurement angle: 0.0125 to 125,000 cd/m2 1° measurement angle: 0.0005 to 5,000 cd/m2 |
||
Accuracy |
Luminance |
±2% |
|
Chromaticity (1° measurement angle) |
x,y: ±0.002 (0.001 to 0.05 cd/m2) x,y: ±0.0015 (from 0.05 cd/m2) |
||
Repeatability |
Luminance (1°C measurement angle) |
1.5% (0.0005 to 0.001 cd/m2) 0.7% (0.001 to 0.003 cd/m2) 0.25% (0.003 to 0.05 cd/m2) 0.15% (0.05 to 5,000 cd/m2) |
|
Chromaticity (1° measurement angle) |
x: 0.003 y: 0.0035 (0.001 to 0.003 cd/m2) x: 0.001 y: 0.0015 (0.003 to 0.1 cd/m2) x: 0.0006 y: 0.0006 (0.1 to 0.2 cd/m2) x: 0.0004 y: 0.0004 (0.2 to 5,000 cd/m2) |
||
Polarization error |
0.1°, 0.2° measurement angle: 3% or less 1° measurement angle: 2% or less |
||
Measurement time |
Approx. 2 to 243 sec. |
||
Color space mode |
Lv x y, Lv u’ v’, Lv TΔuv, XYZ ,spectral waveform, dominant wavelength, excitation purity scotopic luminosity (with CS-S10w Professional) |
||
Interface |
USB 1.1 |
||
Operating temperature/humidity range |
5 to 30°C; Relative humidity 80% or less with no condensation |
||
Power supply |
AC adapter (100 to 240V~, 50/60 Hz) |
||
Dimensions (W × H × D) |
158 × 200 × 300mm (Main unit) Ø70 × 95mm (Lens) |
||
Weight |
6.2kg (including lens) |